Innovative Orthopedic Technologies, LLC: Medical Device Recall in 2021 - (Recall #: Z-0072-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

ARCH Leg Positioning System

Product Classification:

Class II

Date Initiated: September 13, 2021
Date Posted: October 13, 2021
Recall Number: Z-0072-2022
Event ID: 88636
Reason for Recall:

Due to the holes not being bored to the specific diameter, the locking pins are unable to engage under a load.

Status: Terminated
Product Quantity: 10 units
Code Information:

Model: 6813010 Serial Numbers: 6813010.0066; 6813010.0067; 6813010.0069; 6813010.0070; 6813010.0071; 6813010.0072; 6813010.0073; 6813010.0074; 6813010.0075; 6813010.0076;

Distribution Pattern:

U.S. Nationwide distribution in the states of AL, AZ, FL, MI, MN, OK, SD, and TX.

Voluntary or Mandated:

Voluntary: Firm initiated