INNOVATIVE TOMOGRAPHY PRODUCT GMBH: Medical Device Recall in 2021 - (Recall #: Z-2163-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Coaxial Interventional Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. KIR 23/05, Length 50 mm, Diameter 23 G (0.6 mm); Article no. KIR 23/05N, Length 50 mm, Diameter 23 G (0.6 mm); Article no. KIR 23/07, Length 75 mm, Diameter 23 G (0.6 mm); Article no. KIR 23/07N, Length 70 mm, Diameter 23 G (0.6 mm); Article no. KIR 23/10, Length 100 mm, Diameter 23G (0,6mm); Article no. KIR 23/15, Length 150 mm, Diameter 23 G (0.6 mm); Article no. KIR 21/10, Length 100 mm, Diameter 21 G (0.8 mm); Article no. KIR 21/12, Length 120 mm, Diameter 21 G (0.8 mm); Article no. KIR 21/15, Length 150 mm, Diameter 21 G (0.8 mm); Article no. KIR 21/20, Length 200 mm, Diameter 21 G (0.8 mm); Article no. KIR 17/15, Length 150 mm, Diameter 17 G (1.40mm) Article no. KIR 17/15:45, Length 150 mm, Diameter 17 G (1.40mm) *NOT DISTRIBUTED WITHIN THE US
Class II
Sterility Assurance; Coaxial Interventional Needle labeled as sterile may not have been adequately sterilized
Affected Lot Numbers: KIR 23/05: LOT: 2516, 5016, 1417, 0718, 3318, 4318, 3819, 1520; KIR 23/05N: LOT: 0717, 5117, 0518, 4318, 1520; KIR 23/07: LOT: 2516, 4216, 3617, 0718, 3318, 4318, 2119, 3819, 1520, 3320, 4420; KIR 23/07N: LOT: 2916, 4516, 5016, 0717, 1618, 4318, 3819, 1520, 5220; KIR 23/10: LOT: 2016, 2916, 4016, 5016, 0417, 0817, 0917, 1417, 2117A, 2117B, 3917, 4217, 4317, 4517,1618, 4118, 3318, 4818, 4318, 0319, 3319, 4119, 4619, 0520, 1620, 4420, 5220 KIR 23/15: LOT: 1417, 0418, 3319, 2420; KIR 21/10: LOT: 2916, 4016, 1417, 2117A, 2117B, 0418, 4318, 0719, 2119, 3519, 4619,1020, 1620, 5220; KIR 21/12: LOT: 2116, 0717, 1618, 2119, 4619, 1020, 3820; KIR 21/15: LOT: 3416, 0717, 1417, 0418, 3318, 4318, 3819, 1020, 3820; KIR 21/20: LOT: 3617, 1020; KIR17/15: LOT: 2417, 5117, 0918; KIR17/15:45: LOT: 3616, 4216, 2317, 5117, 0918, 3919, 2720;
US Nationwide distribution in the state of CA.
Voluntary: Firm initiated