Inpeco S.A.: Medical Device Recall in 2016 - (Recall #: Z-0587-2017)
See the recall detail below. You can also see other recalls from the same firm in 2016.
FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Class II
Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.
Shucko Socket with P/N: 693822160, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 3 meters; Shucko Socket with P/N: 693822170, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 5 meters; Shucko Socket with P/N: 693822190, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 7 meters; Shucko Socket with P/N: 0C00005640, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 2 meters. The Serial number of the impacted products globally distributed are the following: ACP.009, ACP.024, ACP.039, ACP.054, ACP.069, ACP.084, ACP.100, ACP.115 , ACP.010, ACP.025, ACP.040, ACP.055, ACP.070, ACP.085, ACP.101, ACP.116, ACP.011, ACP.026, ACP.041, ACP.056, ACP.071, ACP.086, ACP.102, ACP.117, ACP.012, ACP.027, ACP.042, ACP.057, ACP.072,ACP.087, ACP.103, ACP.118 , ACP.013, ACP.028, ACP.043, ACP.058, ACP.073, ACP.088, ACP.104, ACP.119, ACP.014, ACP.029, ACP.044, ACP.059, ACP.074, ACP.089, ACP.105, ACP.120, ACP.015, ACP.030, ACP.045, ACP.060, ACP.075, ACP.090, ACP.110, ACP.121, ACP.016, ACP.031, ACP.046, ACP.061, ACP.076, ACP.091, ACP.107, ACP.122 ,ACP.017, ACP.032, ACP.047, ACP.062, ACP.077, ACP.092, ACP.108, ACP.123 ,ACP.018, ACP.033, ACP.048, ACP.063, ACP.078, ACP.093, ACP.109, ACP.124, ACP.019, ACP.034, ACP.049, ACP.064, ACP.079, ACP.094, ACP.110, ACP.125, ACP.020, ACP.035, ACP.050, ACP.065, ACP.080, ACP.096, ACP.111, ACP.126 ,ACP.021, ACP.036, ACP.051, ACP.066, ACP.081, ACP.097, ACP.112, ACP.127 , ACP.022, ACP.037, ACP.052, ACP.067, ACP.082, ACP.098, ACP.113, ACP.128, ACP.023, ACP.038, ACP.053, ACP.068, ACP.083, ACP.099, ACP.114, ACP.129, FLX.103, FLX.0120, FLX.0131, FLX.0137, FLX.0143, FLX.0161, FLX.0166, FLX.0179, FLX.0104, FLX.0124, FLX.0134, FLX.0140, FLX.0152, FLX.0163, FLX.0175, FLX.0108, FLX.0125, FLX.0135, FLX.0141 FLX.0155, FLX.0164, FLX.0176, FLX.0115, FLX.0126, FLX.0136, FLX.0142, FLX.0158, FLX.0165, FLX.0178, AP2.0001, AP2.0004, AP2.0006, AP2.0008, AP2.0010, AP2.0012, AP2.0014, AP2.0003, AP2.0002, AP2.0005, AP2.0007, AP2.0009, AP2.0011, AP2.0013, AP2.0015, FLX.0008, FLX.0133, FLX.0149, FLX.0153, FLX.0159, FLX.0167, FLX.0170, FLX.0174, FLX.0069, FLX.0139, FLX.0150 FLX.0156, FLX.0162, FLX.0168, FLX.0172, FLX.0179, FLX.0132, FLX.0147, FLX.0151, FLX.0157, FLX.0166, FLX.0169, FLX.0173 ,FLX.0121, FLX.0138, FLX.0144, FLX.0145, FLX.0146, FLX.0148, FLX.0176 and FLX.0181.
Worldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain.
Voluntary: Firm initiated