Integra LifeSciences Corp.: Medical Device Recall in 2014 - (Recall #: Z-0197-2015)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.
Class II
A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
Lot Numbers PM0278 and PM0277
Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
Voluntary: Firm initiated