Integra LifeSciences Corp.: Medical Device Recall in 2019 - (Recall #: Z-0613-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Product Classification:
Class II
Date Initiated: October 21, 2019
Date Posted: December 18, 2019
Recall Number: Z-0613-2020
Event ID: 84216
Reason for Recall:
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
Status: Terminated
Product Quantity: 820
Code Information:
US Ops Manual P/N 60905769
Distribution Pattern:
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Voluntary or Mandated:
Voluntary: Firm initiated