Integrity Implants Inc.: Medical Device Recall in 2022 - (Recall #: Z-0419-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 45mm; o. 7.5mm x 50mm; p. 7.5mm x 55mm; q. 7.5mm x 60mm; r. 8.5mm x 35mm; s. 8.5mm x 40mm; t. 8.5mm x 45mm; u. 8.5mm x 50mm; v.8.5mm x 55mm; w. 8.5mm x 60mm; Used in spinal fixation surgery.

Product Classification:

Class II

Date Initiated: January 27, 2022
Date Posted: December 14, 2022
Recall Number: Z-0419-2023
Event ID: 91070
Reason for Recall:

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Status: Terminated
Product Quantity: 1,702 screws
Code Information:

a. 5.5mm x 35mm, Part Number: LS-N6SG5535, UDI: (01)00818613022568(10)02210139, lot Number: 2210139. b. 5.5mm x 40mm, Part Number: LS-N6SG5540, UDI: (01)00818613022575(10)02210140, Lot Number: 2210140. c. 5.5mm x 45mm, Part Number: LS-N6SG5545, UDI: ((01)00818613022582(10)02210141, Lot Number: 2210141. d. 5.5mm x 50mm, Part Number: LS-N6SG5550, UDI: (01)00818613022599(10)02210142, Lot Number: 2210142. e. 5.5mm x 55mm, Part Number: LS-N6SG5555, UDI: (01)00818613022605(10)02210143, Lot Number: 2210143. f. 6.5mm x 35mm, Part Number: LS-N6SG6535, UDI: (01)00818613022643(10)02210144, Lot Number: 2210144. g. 6.5mm x 40mm, Part Number: LS-N6SG6540, UDI: (01)00818613022650(10)02210145, Lot Number: 2210145. h. 6.5mm x 45mm, Part Number: LS-N6SG6545, UDI: (01)00818613022667(10)02210146, Lot Number: 2210146, UDI: (01)00818613022667(10)02210146-1, Lot Number: 02210146-1. i. 6.5mm x 50mm, Part Number: LS-N6SG6550, UDI: (01)00818613022674(10)02210147, Lot Number: 2210147. j. 6.5mm x 55mm, Part Number: LS-N6SG6555, UDI: (01)00818613022681(10)02210148, Lot Number: 2210148. k. 6.5mm x 60mm, Part Number: LS-N6SG6560, UDI: (01)00818613022698(10)02210149, Lot Number: 2210149. l. 7.5mm x 35mm, Part Number: LS-N6SG7535, UDI: (01)00818613022711(10)02210150, Lot Number: 2210150. m. 7.5mm x 40mm, Part Number: LS-N6SG7540, UDI: (01)00818613022735(10)02210152, Lot Number: 2210151. n. 7.5mm x 45mm, Part Number: LS-N6SG7545, UDI: (01)00818613027891(10)02210131, Lot Number: 2210152. o. 7.5mm x 50mm, Part Number: LS-N6SG7550, UDI: (01)00818613022742(10)02210153, Lot Number: 2210153. p. 7.5mm x 55mm, Part Number: LS-N6SG7555, UDI: (01)00818613022759(10)02210154, Lot Number: 2210154. q. 7.5mm x 60mm, Part Number: LS-N6SG7560, UDI: (01)00818613022766(10)02210155, Lot Number: 2210155. r. 8.5mm x 35mm, Part Number: LS-N6SG8535, UDI: (01)00818613022780(10)02210156, Lot Number: 2210156. s. 8.5mm x 40mm, Part Number: LS-N6SG8540, UDI: (01)00818613022797(10)02210157, Lot Number: 2210157. t. 8.5mm x 45mm, Part Number: LS-N6SG8545, UDI: (01)00818613022803(10)02210158, Lot Number: 2210158. u. 8.5mm x 50mm, Part Number: LS-N6SG8550, UDI: (01)00818613022810(10)02210159, Lot Number: 2210159. v.8.5mm x 55mm, Part Number: LS-N6SG8555, UDI: (01)00818613022827(10)02210160, Lot Number: 2210160. w. 8.5mm x 60mm, Part Number: LS-N6SG8560, UDI: (01)00818613022834(10)02210161, Lot Number: 2210161.

Distribution Pattern:

Nationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.

Voluntary or Mandated:

Voluntary: Firm initiated