Inter-Med Llc: Medical Device Recall in 2019 - (Recall #: Z-0012-2020)
See the recall detail below. You can also see other recalls from the same firm in 2019.
Vista Dental Products Sodium Hypochlorite - 6%, 10 - 12cc Pre-filled Syringes, REF 502300
Class II
The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
Lot/Work Order Numbers: 2018-1348 2018-1585 2018-1586 2018-1852 20182106, UDI (01)10818207020465 20182302, UDI (01)10818207020465 20182458, UDI (01)10818207020465 20190725, UDI (01)10818207020465 20190835, UDI (01)10818207020465
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Voluntary: Firm initiated