Inter-Med Llc: Medical Device Recall in 2019 - (Recall #: Z-0015-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Vista Dental Products 3% Sodium Hypochlorite, 10- 3cc Pre-filled Syringes, REF 502355

Product Classification:

Class II

Date Initiated: June 27, 2019
Date Posted: October 9, 2019
Recall Number: Z-0015-2020
Event ID: 83435
Reason for Recall:

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Status: Terminated
Product Quantity: 445 units
Code Information:

UDI (01)10818207020410 Lot/Work Order Numbers: 20182545 20182665 20190198 20190432 20190724 20190953 20191314

Distribution Pattern:

US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Voluntary or Mandated:

Voluntary: Firm initiated