International Biomedical, Ltd.: Medical Device Recall in 2013 - (Recall #: Z-1531-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).
Class II
International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to
Model/Part Number: 209-2113, or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, 4489
Nationwide Distribution.
Voluntary: Firm initiated