Intuitive Surgical, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1431-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.

Product Classification:

Class II

Date Initiated: May 8, 2013
Date Posted: June 5, 2013
Recall Number: Z-1431-2013
Event ID: 65194
Reason for Recall:

Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.

Status: Terminated
Product Quantity: 697 units - all products
Code Information:

Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M10121115 M10121123 M10121130 M10121206 M10121210 M10130124 M10130131 M10130205 M10130304 M10130319 M10130321

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated