Intuitive Surgical, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1432-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
Class II
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Model number: 400172-08; Lot numbers: M10120815 M10120817 M10120831 M10120913 M10120918 M10121008 M10121024 M10121120 M10121123 M10121203 M10121210 M10121215 M10130220 M10130222 M10130311 M10130315 M10130327 M10130411.
Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.
Voluntary: Firm initiated