Intuitive Surgical, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1433-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
8mm Fenestrated Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
Class II
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Model number: 400205-04; Lot numbers: M10120816 M10120820 M10120903 M10120910 M10120912 M10120921 M10121008 M10121024 M10121120 M10121130 M10121206 M10121213 M10121215 M10130124 M10130131 M10130213 M10130218 M10130222 M10130313 M10130319 M10130402.
Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.
Voluntary: Firm initiated