Product Description:

EndoWrist Stapler 45 instrument for the da Vinci Si System Model Number: IS3000 Product Usage: For use with da Vinci Si System. The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, transoral otolaryngology surgical procedures restricted to benign and malignant tumors classified as TI and T2, general thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use (except for transoral otolaryngology surgical procedures). It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Product Classification:

Class II

Date Initiated: September 18, 2014
Date Posted: October 22, 2014
Recall Number: Z-0074-2015
Event ID: 69283
Reason for Recall:

Intuitive is issuing an urgent product notice to all users of the EndoWrist Stapler 45 to discontinue use due to reports of inability to remove the stapler from tissue.

Status: Terminated
Product Quantity: 1,034 Devices
Code Information:

EndoWrist Stapler 45 instrument, Model Number: IS3000 Part Number: PN 410298-06, -07, -08, -09 All Lots.

Distribution Pattern:

US Nationwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated