Intuitive Surgical, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0445-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
8MM, SMALL GRASPING RETRACTOR ; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
Product Classification:
Class II
Date Initiated: November 3, 2014
Date Posted: December 3, 2014
Recall Number: Z-0445-2015
Event ID: 69638
Reason for Recall:
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
Status: Terminated
Product Quantity: 509 total units, all affected devices
Code Information:
Part number; 470318-04 ; All 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings.
Distribution Pattern:
US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
Voluntary or Mandated:
Voluntary: Firm initiated