Intuitive Surgical, Inc.: Medical Device Recall in 2015 - (Recall #: Z-2751-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Medium -Large Clip Applier; used with the da Vinci Si Surgical System (IS3000); Model numbers: 428053-12, 428053-15. The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation and electrocautery during single incision laparoscopic cholecystectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, scissors, suction irrigators, monopolar cautery, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, flexible blunt obturators, and the 5 mm Single-Site Port.

Product Classification:

Class II

Date Initiated: August 8, 2015
Date Posted: September 16, 2015
Recall Number: Z-2751-2015
Event ID: 72036
Reason for Recall:

In certain circumstances involving select da Vinci Si Single-Site instruments, the jaws of grip actuated instruments may become fixed in a closed position.

Status: Terminated
Product Quantity: 17617 total units
Code Information:

Model numbers: 428053-12, 428053-15. ; all lots

Distribution Pattern:

Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, Canada, Chile, China, Cyprus, Denmark, France, Germany, Greece, Israel, Italy, Monaco, Netherlands, Panama, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated