Intuitive Surgical, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2381-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
da Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.
Class II
Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.
UDI: 00886874112823. Device expiration: 10/14/2017 02/21/2020
U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Voluntary: Firm initiated