Intuitive Surgical, Inc.: Medical Device Recall in 2020 - (Recall #: Z-1105-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298
Class II
This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
Serial Number/Manufacturing Lot: N11181003, N11181004
US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming OUS:None
Voluntary: Firm initiated