Intuitive Surgical, Inc.: Medical Device Recall in 2024 - (Recall #: Z-2720-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Ion Endoluminal System, REF: 380748-65
Product Classification:
Class II
Date Initiated: July 12, 2024
Date Posted: September 4, 2024
Recall Number: Z-2720-2024
Event ID: 95046
Reason for Recall:
There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
Status: Ongoing
Product Quantity: 3 units
Code Information:
UDI: 00886874116234/ System Serial Number/System Cart Serial Number: EN1230/10591271; EN1241/10591273; EN1231/10583014
Distribution Pattern:
US Nationwide distribution in the states of AZ and CA.
Voluntary or Mandated:
Voluntary: Firm initiated