Invacare Corporation: Medical Device Recall in 2014 - (Recall #: Z-2091-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

MK5 NX Controller--Wheelchair controller

Product Classification:

Class II

Date Initiated: June 24, 2014
Date Posted: July 30, 2014
Recall Number: Z-2091-2014
Event ID: 68622
Reason for Recall:

M41 Powered Wheelchairs received through Invacare Service Parts (ISP) have controllers that were programmed in a way this is not aligned to standard factory settings. The controller during different settings such as max forward speed, max forward acceleration, max reverse speed and max turning speed can result in momentary loss of control.

Status: Terminated
Product Quantity: 819
Code Information:

P/N 1143839

Distribution Pattern:

Nationwide Distriution: CA, SC, IL, UT, TN, KS, HI, NJ, LA, TX,FL, IN, AL, AZ, MN, WI, MN, MI, AR,MA, GA, MS, NJ, NM, OK, ME, PA, NY, NC, OK, MO, WA, IA, OR,KY, MD, NE, ID, VA, WA, ND, MT, OH, NH, CT. Also shipped to Canada

Voluntary or Mandated:

Voluntary: Firm initiated