Ion Beam Applications S.A.: Medical Device Recall in 2017 - (Recall #: Z-2592-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Proteus 235
Class II
It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.
Serial Numbers: PAT.000,PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.111, PAT.112, PAT.113, PAT.114, SAT.115, SAT.116, SAT.117, SAT.119, SAT.120, SAT.122, SAT.123, SAT.126
Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
Voluntary: Firm initiated