Ion Beam Applications S.A.: Medical Device Recall in 2021 - (Recall #: Z-2155-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Proteus 235-Proton Therapy System

Product Classification:

Class II

Date Initiated: May 26, 2021
Date Posted: August 4, 2021
Recall Number: Z-2155-2021
Event ID: 88196
Reason for Recall:

if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment field may be delivered with an error in position

Status: Ongoing
Product Quantity: 5 US; 8 OUS
Code Information:

Serial Numbers: PAT.107 (EU), PAT.108 (US), PAT.109 (US), PAT.115 (EU), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.125 (IN), SAT.126 (US), SAT.132 (EU), SAT.133 (US).

Distribution Pattern:

FL, PA, TN, TX, VA Foreign: Germany, Italy, Russia, Sweden , India, Netherlands

Voluntary or Mandated:

Voluntary: Firm initiated