Ion Beam Applications S.A.: Medical Device Recall in 2022 - (Recall #: Z-0747-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235
Class II
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
PAT.003 (KR), PAT.006 (US), PAT.107 (EU), PAT.108 (US), PAT109 (US), PAT.110 (US), PAT.111 (EU), PAT.112 (US), PAT.113 (US), PAT.114 (EU), PAT.115 (EU), SAT.116 (US), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (EU), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (EU), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.109 (UK), SBF.110 (UK), SBF.112 (EU), SBF.113 (US), SBF.115 (UK), SBF.117 (EU), SBF.128 (US). UDI: (01)05404013801138
IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,
Voluntary: Firm initiated