Jjgc Industria E Comercio De Materials Dentarios Sa: Medical Device Recall in 2021 - (Recall #: Z-0203-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
GM Helix Acqua Implant 3.5x10, Article No. 140.944
Product Classification:
Class II
Date Initiated: October 11, 2021
Date Posted: November 10, 2021
Recall Number: Z-0203-2022
Event ID: 88917
Reason for Recall:
3.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is installed, it could result in damage to the nerve or adjacent bone structures/cavities.
Status: Terminated
Product Quantity: 567 (US)
Code Information:
UDI 7899878024897 Lots GNW15 and GRE82
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated