Johnson & Johnson Surgical Vision, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1006-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

FUSION Dual Pump Packs, REF: OPO73

Product Classification:

Class II

Date Initiated: December 7, 2022
Date Posted: February 1, 2023
Recall Number: Z-1006-2023
Event ID: 91290
Reason for Recall:

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Status: Ongoing
Product Quantity: 7644
Code Information:

UDI-DI/Lots: 15050474602080/60373238, 60394037, 60308197

Distribution Pattern:

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated