Johnson & Johnson Surgical Vision, Inc.: Medical Device Recall in 2023 - (Recall #: Z-2247-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Product Classification:

Class II

Date Initiated: May 30, 2023
Date Posted: August 9, 2023
Recall Number: Z-2247-2023
Event ID: 92539
Reason for Recall:

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Status: Ongoing
Product Quantity: 148,941
Code Information:

All lots within their expiration date as of 5/29/2023. Part Number/UDI-DI/Lot: VRT-AI/05050474700901/60298931, 60302489, 60302939, 60302941, 60302953, 60302954, 60304170, 60304175, 60304176, 60305660, 60305661, 60305662, 60305665, 60306934, 60306935, 60306936, 60306940, 60306942, 60308193, 60309850, 60309844, 60309845, 60309853, 60314642, 60314676, 60314677, 60316112, 60351692, 60372490, 60372491, 60374175, 60374945, 60376930, 60378840, 60378842, 60378844, 60400795, 60401984, 60413115, 60413116, 60413117, 60414191, 60415077, 60415078, 60416304, 60420003, 60421952, 60425389, 60426314, 60426315, 60428217, 60428216, 60429442, 60433653, 60433654, 60433655, 60435921, 60435922, 60435923, 60437988, 60437989, 60440239, 60440240, 60442450, 60442451, 60442452, 60442453, 60442454, 60444535, 60446059, 60448847, 60448856, 60448857, 60376931, 60379976, 60429443. VRT-AF/05050474700895/60304172, 60304173, 60304174, 60305657, 60305658, 60305659, 60306930, 60306931, 60306932, 60306933, 60308189, 60308190, 60309848, 60309847, 60314640, 60314641, 60314674, 60314675, 60378839, 60420004, 60446054, 60446055

Distribution Pattern:

Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Korea¿(South), Kuwait, Lebanon, Malaysia, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated