Kalila Medical: Medical Device Recall in 2018 - (Recall #: Z-0563-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.
Product Classification:
Class II
Date Initiated: December 14, 2017
Date Posted: February 21, 2018
Recall Number: Z-0563-2018
Event ID: 78824
Reason for Recall:
Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.
Status: Terminated
Product Quantity: 99 sheaths
Code Information:
Lot #0559, exp. 10/9/2018, and Lot #0571, exp. 10/19/2018.
Distribution Pattern:
Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.
Voluntary or Mandated:
Voluntary: Firm initiated