Ameritus Medical Products ENFit Polyurethane and Silicone Feeding Tubes. Sterile and single use only. Catalog number ENF-Y60P-50 (6 fr. 50 cm ENFit Enteral Feeding Tube with Y-Port).
Class II
Feeding tube plastic tethered closure cap may possibly fracture or break. Possible health hazards: (1) Enteral feeding and medication administration could be possibly delayed, until replacement tube is placed into patient (2) Patient could be unnecessarily overexposed to radiation if multiple x-rays are needed to confirm placement of a new tube, if x-ray is used as the placement confirmation method.
Lot Number: KS1810008
US: CA, IA, and VA
Voluntary: Firm initiated