Kerr Corporation: Medical Device Recall in 2012 - (Recall #: Z-0402-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.

Product Classification:

Class II

Date Initiated: November 17, 2011
Date Posted: November 28, 2012
Recall Number: Z-0402-2013
Event ID: 63512
Reason for Recall:

The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.

Status: Terminated
Product Quantity: 1330 units
Code Information:

Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System has no expiration date. The affected units were produced and distributed since February 2011.

Distribution Pattern:

Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.

Voluntary or Mandated:

Voluntary: Firm initiated