King Systems Corp.: Medical Device Recall in 2012 - (Recall #: Z-0424-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg Product Usage: Laryngeal airway
Class II
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
REF: LAD-705, Lot #K120413
US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Voluntary: Firm initiated