L3 Security & Detection: Medical Device Recall in 2018 - (Recall #: Z-0270-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.

Product Classification:

Class II

Date Initiated: October 27, 2017
Date Posted: January 3, 2018
Recall Number: Z-0270-2018
Event ID: 78658
Reason for Recall:

L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel. This a gap in x-ray shielding can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).

Status: Terminated
Product Quantity: 471
Code Information:

Model Number : Examiner 3DX 6000 cabinet x-ray system

Distribution Pattern:

USA (nationwide) Distribution.  Owned and operated by the U.S. Transportation Security Administration. Location of individual products is law enforcement sensitive. TSA HQ is located in Crystal City, Virginia.

Voluntary or Mandated:

FDA Mandated