Landauer: Medical Device Recall in 2023 - (Recall #: Z-2542-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
nanoDot D2DNS, Model Numbers: a) 03055-1MO; b) 03055-3MO; c) 03055-OTO; d) 03055-SMO; radiation monitoring dosimeter used with the microSTAR readers
Product Classification:
Class II
Date Initiated: July 12, 2023
Date Posted: September 20, 2023
Recall Number: Z-2542-2023
Event ID: 92647
Reason for Recall:
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Status: Completed
Product Quantity: 60 units
Code Information:
UDI/DI 0860003399903, all batch numbers
Distribution Pattern:
Worldwide
Voluntary or Mandated:
Voluntary: Firm initiated