LEICA BIOSYSTEMS NUSSLOCH GMBH: Medical Device Recall in 2024 - (Recall #: Z-0085-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples
Product Classification:
Class II
Date Initiated: September 4, 2024
Date Posted: October 23, 2024
Recall Number: Z-0085-2025
Event ID: 95323
Reason for Recall:
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Status: Ongoing
Product Quantity: 1 unit
Code Information:
UDI/DI 04049188066237, Serial Number 990
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated