LEICA BIOSYSTEMS NUSSLOCH GMBH: Medical Device Recall in 2024 - (Recall #: Z-0086-2025)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples
Class II
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
a) Material Number 1491510C002, UDI/DI (01)04049188002945, Serial Numbers: 5359022006, 5367012006, 5913, 5996, 5065, 5625022007, 5467072006, 5901, 5108122004, 5330, 5269, 5667; b) Material Number 1491510C003, UDI/DI (01)04049188002891. Serial Numbers: 5537
US Nationwide distribution.
Voluntary: Firm initiated