LEICA BIOSYSTEMS NUSSLOCH GMBH: Medical Device Recall in 2024 - (Recall #: Z-0090-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples

Product Classification:

Class II

Date Initiated: September 4, 2024
Date Posted: October 23, 2024
Recall Number: Z-0090-2025
Event ID: 95323
Reason for Recall:

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Status: Ongoing
Product Quantity: 187 units
Code Information:

a) Material Number 14047033518, UDI/DI 04049188065599, Serial Numbers: 5013012008, 3151052002, 3559082003; b) Material Number 14903050S01, UDI/DI 04049188003270, Serial Numbers: 3641, 4119012005, 5212102008, 4002102004, 3944092004, 3137042002, 3683122003, 5119, 4125022005, 4171032005, 3954092004, 2636102000, 3847052004, 4387122005, 4585092006, 4734022007, 3951092004, 3342112002, 5208082008, 4575092006, 4521052006, 2529052000, 2605, 4384102005, 4538062006, 4807, 4269062005, 2700122000, 3157052002, 4094012005, 4503042006, 3564082003, 4654, 2857072001, 3874062004, 2418022000, 3417012003, 4572082006, 3349, 3748022004, 4850072007, 4948102007, 3588092003, 3136042002, 4415122005, 4454022006, 3240072002, 5104052008, 3009012002, 4142, 4014112004, 2421022000, 3376, 3901072004, 3163052002, 3917072004, 2781042001, 4223052005, 4301092005, 5131062008, 4724022007, 2676112000, 3589092003, 4145, 4569082006, 3508, 2533062000, 4311092005, 4317092005, 4335092005, 4795052007, 5207, 4573082006, 5152072008, 5159082008, 3277092002, 3662112003, 3078032002, 3956092004, 4708022007, 4127022005, 3677122003, 4845, 3260092002, 3394, 3037012002, 4698012007, 5011012008, 2891082001, 3540072003, 3869062004, 2525052000, 4514052006, 4299072005, 3915042004, 5083, 3307102002, 5155082008, 3227072002, 3952092004, 5007012008, 5088, 2392121999, 3162052002, 3346112002, 4394, 3738022004, 3758032004, 5015012008, 4735022007, 2840062001, 3219072002, 3755022004, 4310092005, 4445012006, 5004122007, 4819062007, 3759032004, 4851, 3093032002, 3945, 3434022003, 4270062005, 4755, 4169032005, 4004102004, 2583072000, 2791/4.2001, 3396, 3824, 4655112006, 3044022002, 3832, 4853082007, 5197102008, 4193032005, 4242, 4764032007, 4781, 5018012008, 2580072000, 4952102007, 2774032001, 4797, 5171082008, 5082042008, 3548072003, 3526062003, 4774, 4320092005, 4124022005, 2943102001, 3375122002, 4801052007, 5224112008; c) Material Number 14903050S02, UDI/DI (01)04049188003201, Serial Numbers: 2624052001, 4802052007, 2633092000, 3187062002, 3908072004, 3459032003, 2890082001, 3783, 3609092003, 2627092000, 3344112002, 3795042004, 2853072001, 3401102003, 4393, 4391102005; d) Material Number 14903050S03, UDI/DI 04049188003188, Serial Numbers: 4926092007, 4644112006, 3004012002, 5226, 4958, 4518, 3644, 5070

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated