Leica Microsystems, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1698-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Class II
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.
Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06
Nationwide Distribution including AZ, CA, FL, and NY.
Voluntary: Firm initiated