Leica Microsystems, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1710-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro diagnostic use.
Class II
Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.
Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: 6017936, Expires 2014-08; 6015073, Expires 2014-03
Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom
Voluntary: Firm initiated