Leica Microsystems, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0475-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system

Product Classification:

Class II

Date Initiated: November 21, 2022
Date Posted: December 21, 2022
Recall Number: Z-0475-2023
Event ID: 91200
Reason for Recall:

There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

Status: Ongoing
Product Quantity: 43 units
Code Information:

Serial Numbers: 20522001, 20722001, 21221001, 40222001, 40722001, 70722001, 80522002, 110722001, 150422001, 180322001, 211021001, 260722001, 290422001, 300122001, 270522001. ***EXPANSION 09/08/2023, Serial Numbers: 271222001, 100423002, 31022001, 180623001, 71222001, 210523002, 270922002, 120623002, 100123001, 130623001, 290922002, 300922002, 210223001, 220223001, 251122001, 281122001, 20623002, 60922001, 200822001, 40523001, 40523002, 270922001, 61122002, 20423001, 300523002, 190822002, 300922001, 280922001.

Distribution Pattern:

US Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated