Leica Microsystems, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0476-2023)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
Class II
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
***EXPANSION 09/08/2023 UPDATE a. Premium Part Number 10448950; Serial Numbers: 160222001, 280622001, 50422001, 50422002, 91221001, 180622001, 170622001, 61221001, 130922001, 120623001, 150623001, 81022001, 200623001, 51022001, 61022001. b. Standard Part Number 10448976; Serial Numbers: 120322001, 300122001, 40222002, 301221001, 40122001, 140522001, 151221002, 81221002, 100222002, 300122002, 110322001, 50322001, 90622001, 120422001, 90522001, 100522001, 250122001, 100222001, 171022001, 80922001, 70623001, 30922001, 91022001.
US Nationwide
Voluntary: Firm initiated