Leica Microsystems, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0477-2023)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system
Class II
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
Serial Numbers: 10622001, 10622002, 10822001, 20322002, 20522001, 20622001, 30122001, 30322001, 30522001, 30622002, 30722001, 31221001, 40422001, 41221001, 50122001, 60122001, 60522001, 60622001, 70122001, 70322001, 70422001, 70622001, 80422001, 80422002, 80722001, 80822001, 90122001, 90622001, 91221001, 100122001, 100122002, 100322002, 100622001, 100622002, 110222001, 110522001, 110822001, 111221001, 120222001, 120522001, 130122001, 130522001, 130622001, 140122001, 140222001, 140422001, 140522001, 141221001, 151221001, 160122001, 160322001, 160622001, 160622002, 170222001, 170522001, 190322001, 190622001, 200122001, 200522001, 200722001, 210422001, 210622001, 230222001, 230622001, 240122001, 240522001, 260522001, 260722001, 260722002, 261121001, 261221001, 270122001, 270422001, 270422002, 270522001, 270622001, 271121001, 280222002, 280622001, 290522001, 290622001, 291221001, 300322001, 300622002, 301121001, 301221001, 310322001, 310522001. ***EXPANSION 09/08/2023, Serial Numbers: 220523002, 220822001, 131222001, 251122001, 271122001, 100223001, 220423001, 300822001, 30323001, 41222001, 51222001, 91222002, 81222002, 301122001, 180822001, 71222001, 180722001, 220523001, 300523002, 80623001, 30523001, 11122001, 31122001, 81122001, 91222001, 30723001, 211122001, 121222001, 310523001, 40423001, 290323001, 300323001, 141222001, 141222002, 310822001, 40623002, 300922001, 200223001, 90223001, 180523001, 280523001, 240423001, 10623001, 191122001, 61222001, 90623001, 190423001, 70223001, 90223002, 40723001, 130623001, 180423001, 230922001, 260922001, 251022001, 291022001, 200323001, 261022001, 140523001, 160523001.
US Nationwide
Voluntary: Firm initiated