Leica Microsystems, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1851-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950
Class II
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
a. 10448976, Serial Numbers: 120322001, 300122001, 170921001, 301221001, 40122001, 140522001, 151221002, 301021002, 81221002, 100222002, 40222002, 220821001, 300122002, 110322001, 50322001, 180921001, 260821001, 90622001, 120422001, 100522001, 90522001, 210821001, 250122001, 301021001, 100222001; b. 10448950, Serial Numbers: 160222001, 90721001, 50422001, 50422002, 91221001, 160921001, 61221001
US Nationwide distribution.
Voluntary: Firm initiated