LeMaitre Vascular, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1751-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

Product Classification:

Class II

Date Initiated: April 11, 2025
Date Posted: May 21, 2025
Recall Number: Z-1751-2025
Event ID: 96716
Reason for Recall:

During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

Status: Ongoing
Product Quantity: 65,977 units
Code Information:

UDI-DI: 00840663101535, 00840663101542. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56. Lots not expired with the following prefix: POC, QPO, XPO.

Distribution Pattern:

Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECHIA, DENMARK, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORTH MACEDONIA, NORWAY, OMAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKIYE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.

Voluntary or Mandated:

Voluntary: Firm initiated