Life Technologies Corporation: Medical Device Recall in 2018 - (Recall #: Z-1279-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Oncomine Dx Target Test User Guides and Assay Definition File, Model: A32461; UDI: (01)10190302006071(11)170911(10)1.8(240)A32461 The Oncomine Dx Target Test is a qualitative in vitro diagnostic test that uses targeted high- throughput, parallel-sequencing technology to detect single-nucleotide variants (SNVs) and deletions in 23 genes from DNA and fusions in ROS1 from RNA isolated from formalin-fixed, paraffin-embedded (FFPE) tumor tissue samples from patients with non-small cell lung cancer (NSCLC) using the Ion PGM Dx System. The test is indicated to aid in selecting NSCLC patients for treatment with the targeted therapies listed in Table 1 in accordance with the approved therapeutic product labeling.

Product Classification:

Class II

Date Initiated: December 4, 2017
Date Posted: April 11, 2018
Recall Number: Z-1279-2018
Event ID: 79437
Reason for Recall:

The Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.

Status: Terminated
Product Quantity: 9
Code Information:

Lot Numbers: v1.8 (Rev. C) (REVC20170906) and v1.8 (Rev. C) (REVC27JUN2017) UDI: (01)10190302006071

Distribution Pattern:

US Distribution to the states of: AZ, CA, NC, TX, and VA.

Voluntary or Mandated:

Voluntary: Firm initiated