Life Technologies Holdings Pte Ltd: Medical Device Recall in 2019 - (Recall #: Z-0187-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Torrent Suite Dx Software Version 5.6.4 Catalog Number:A33178 OEM Customer software: Sentosa SQ Suite Software v5.6.20 UDI: (01)10190302005654 (11)000000 (10)5.6.4 (240)A33178

Product Classification:

Class II

Date Initiated: September 26, 2019
Date Posted: October 30, 2019
Recall Number: Z-0187-2020
Event ID: 83786
Reason for Recall:

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Status: Terminated
Product Quantity: total of Ion+vela +chassis =400 units
Code Information:

The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type

Distribution Pattern:

states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

Voluntary or Mandated:

Voluntary: Firm initiated