Lifescan Inc: Medical Device Recall in 2013 - (Recall #: Z-1074-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

Product Classification:

Class I

Date Initiated: March 11, 2013
Date Posted: April 24, 2013
Recall Number: Z-1074-2013
Event ID: 64624
Reason for Recall:

The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.

Status: Terminated
Product Quantity: US 710,413
Code Information:

Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-02; Verio IQ Warranty Kits: 022 275 01, 022 275 02. All lots and serial numbers

Distribution Pattern:

USA Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated