Linvatec Corp. dba ConMed Linvatec: Medical Device Recall in 2013 - (Recall #: Z-1681-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.
Class II
Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
C7122 Lots #'s: 1301224, 1301234, 1301244, 1301304, 1301314, 1302114, 1302154, 1302274, 1302284, 1303144, 1303254, 1304044, 1304234, 1304244, and 1305024
Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.
Voluntary: Firm initiated