LivaNova USA Inc.: Medical Device Recall in 2020 - (Recall #: Z-1498-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.

Product Classification:

Class II

Date Initiated: February 24, 2020
Date Posted: March 25, 2020
Recall Number: Z-1498-2020
Event ID: 85098
Reason for Recall:

The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.

Status: Terminated
Product Quantity: Total Worldwide=9309 systems (2713 system in U.S.)
Code Information:

Device Listing: E572259 Product Code Product Description Affected Serial Number range 16-02-80 Heater-Cooler 3T, 230V 16S10027 16S16874 16-02-81 Heater-Cooler 3T, 240V 16S10743 16S11708 16-02-82 Heater-Cooler 3T, 208V 16S10772 16S16840 16-02-83 Heater-Cooler 3T, 127V 16S11455 16S16509 16-02-85 Heater-Cooler 3T, 120V 16S10958 16S16847 16-02-95 Heater-Cooler 3T, 200V 16S12004 16S16818

Distribution Pattern:

U.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none

Voluntary or Mandated:

Voluntary: Firm initiated