Livongo Health: Medical Device Recall in 2016 - (Recall #: Z-1587-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the self- monitoring of blood glucose by persons with diabetes in home settings to aid in their diabetes management.
Class II
The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid
Model/Reorder Number 3614-001-001 with the following lot numbers 890071 890072, 890073, 890074, 890075, 890076, 890077, 890078 and 890079.
US consignees: Arizona, California, Delaware, Florida, Georgia, Illinois, Kentucky, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia and Washington.
Voluntary: Firm initiated