Luminex Corporation: Medical Device Recall in 2019 - (Recall #: Z-2284-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

ARIES Flu A/B & RSV Assay , REF 50-10020, UDI # 00840487100158

Product Classification:

Class II

Date Initiated: February 20, 2019
Date Posted: August 21, 2019
Recall Number: Z-2284-2019
Event ID: 83470
Reason for Recall:

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

Status: Ongoing
Product Quantity: N/A
Code Information:

Lot numbers and discontinue use by dates for Part Number 50-10020: AA7127 - Immediately; AA7206 - Immediately; AA7312 - 3/5/2019; AA7549 - 3/30/2019; AA7635 - 4/10/2019

Distribution Pattern:

US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated