Luminex Corporation: Medical Device Recall in 2019 - (Recall #: Z-2284-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
ARIES Flu A/B & RSV Assay , REF 50-10020, UDI # 00840487100158
Class II
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Lot numbers and discontinue use by dates for Part Number 50-10020: AA7127 - Immediately; AA7206 - Immediately; AA7312 - 3/5/2019; AA7549 - 3/30/2019; AA7635 - 4/10/2019
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Voluntary: Firm initiated