LW Scientific, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1050-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips
Class II
The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.
Serial Numbers: 6C1803035, 6C1803066, 6C1803121, 6C1803130, 6C1803132, 6C1803133, 6C1803139, 6C1803141, 6C1803142, 6C1803144, 6C1803145, 6C1805217, 6C1805242, 6C1805243, 6C1810375, 6C1810441
Worldwide Distribution - US Nationwide and the countries of Philippines, Honduras, El Salvador, Australia, Guatemala, Canada, Singapore, Australia, Peru, Nicaragua, Trinidad and Tobago, and New Zealand
Voluntary: Firm initiated