Magellan Diagnostics, Inc.: Medical Device Recall in 2021 - (Recall #: Z-1953-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

LeadCare Plus Blood Lead Test Kit Catalog Number 82-0004

Product Classification:

Class I

Date Initiated: May 7, 2021
Date Posted: July 7, 2021
Recall Number: Z-1953-2021
Event ID: 87992
Reason for Recall:

Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.

Status: Terminated
Product Quantity: 320 kits (96 tests/kit=30,720 tests)
Code Information:

Lot Numbers/Expiration Date: 2011MU - 25MAR22. 1st expansion 6/21/2021: 2104MU - 25AUG22; 2108MU - 31MAR22.

Distribution Pattern:

Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.

Voluntary or Mandated:

Voluntary: Firm initiated